Operating Procedure of Maglumi X3


07 January, 2026

Consumables:

CodeNAMESTORAGEUsageFunction
1Reaction module

No sunlight or humidity.

Shelf life: 2 years.

DisposableCLIA Reaction place
2Wash Concentrate

15-30°C, Shelf life: 1 year.

On board stability: 4 weeks.

1 bottle (714mL) + distill water = 10L Wash liquidWash the magnetic microbeads and needles.
3Light Check2-8°C, Shelf life: 1 year.

Avoid bubbles

Special LC Kit

The last reagent position

Check the performance of Pipetting unit and measuring unit
4

Starter 1: Catalyst in 1.5% NaOH

Starter 2: 0.21% hydrogen peroxide solutions

15-30°C, Shelf life: 1 year

On board stability: 4 weeks

Avoid mistakeActivate ABEI
5Tubing Cleaning Solution15-30°C, Shelf life: 1 year.5th, 6th and 7th vials with 13ml respectively.Wash waste tubes

Run Sample

  1. Load the sample in the sample rack.
  2. Scan and edit the sample information in the worklist interface.

Prepare Sample

  1. Collect and deal with samples gently and avoid shaking violently; fibrin, hemolysis, bubbles might affect the results.
  2. Avoid repeated freezing and thawing (< 2 times).
  3. Centrifugation for serum samples: 10-15 min with over 4000r/min.
  4. Use the required sample tube. 